SOP ON OPERATION AND CALIBRATION OF COLONY COUNTER
-
OBJECTIVE:
The purpose of this SOP is to lay down the procedure for Operation and Calibration of Colony Counter.
-
SCOPE:
This SOP is applicable for Operation and Calibration of Colony Counter ID No: —— used for the enumeration of microbial colonies on the agar plates in the microbiological analysis in the microbiology laboratory at a manufacturing facility.
-
RESPONSIBILITY:
- Microbiologist or his designee shall be responsible for Operation and Calibration of Colony Counter.
- A Microbiologist is responsible for related documentation activities.
- Head of Microbiology is responsible for effective implementation of the SOP.
- QA is responsible for review and Approval of the SOP.
-
PROCEDURE:
-
Operation:( Manual Colony Counter )
- Clean the instrument with IPA before the operation start.
- Place the instrument in a clean and dry place. Switch ON the power switch of the instrument
- Ensure that the marker is connected to the instrument.
- Read the digital screen display, it should be “0000” if not then do it by pressing the reset button manually.
- Now place the Petri dish in an inverted position on the instrument platform to count the colonies by marking on each colony.
- The number of colonies will be counted and displayed on the screen.
- After completion of the work microbiologist should enter the details in the Instrument usage log book as per Annexure II.
-
-
Calibration:
- Calibration Frequency: Monthly.
- Attach UNDER CALIBRATION TAG on to the instrument.
- Follow the operating instructions for the operation.
- Press the tip of the marker pen 5 times, 10 times, 15 times, 20 times, 25 times and then check the number displayed on the screen.
- After completion of the calibration microbiologist should enter the details in the Calibration record of Colony counter and in Instrument usage log book.
- Acceptance Criteria: In all the cases the number displayed on the screen should not be differ from No. of a press of marker pen tip.
-
TRAINING:
Trainer: Head of Microbiology
Trainees: Microbiologist
-
DISTRIBUTION:
As Per Requirement
-
ANNEXURE(S):
Annexure- I: Calibration Record of Colony Counter.
Annexure-II: Colony Counter Usage Record.
-