SOP on Analytical Report (AR) Numbering (Example)
Contents
In short: An Analytical Report (AR) number is a unique identifier that links a sample, its test data and its result. This is an illustrative example SOP for QC departments. The number formats shown are an example in-house criterion: no regulation prescribes them, but the records must be traceable and unique.
Why AR numbering matters
GMP does not dictate an AR number format. It requires that laboratory records are complete and traceable to the sample tested:
- 21 CFR 211.194(a) requires laboratory records to include complete data from all tests, including a description of the sample (source, quantity, lot or code number, dates of sampling and receipt), the raw data and calculations, and the signature of the analyst plus a second-person review.
- EU GMP Part I Chapter 6 (in operation since 2014) says test records should include the material or product name, the batch number, references to specifications and test procedures, results with observations and calculations, and dates of testing (clause 6.17), and that sampling is recorded under written procedures (clause 6.11).
A unique AR number is one practical way to meet these expectations. Whatever format you choose, the system must ensure numbers are issued once, in sequence, and cannot be reused.
Objective
To define how Analytical Report (AR) numbers are assigned in the Quality Control department.
Scope
Applies to all samples tested in the QC laboratory at the manufacturing facility: raw materials, packing materials, in-process and finished products, water, cleaning samples and miscellaneous items.
Responsibility
- Analyst / Section Head (or designee): follow this SOP and obtain the AR number from the register.
- Head of Quality (or designee): ensure implementation and periodic review of the registers.
Procedure (example in-house criterion)
Format keys: YY = last two digits of the year; ABCD = running serial number (restarted each year, or as per your SOP); the company name or site code goes first. Issue each number from a controlled register, never reuse it, and record cancelled numbers with a reason.
| Sample type | AR number format | Code meaning |
|---|---|---|
| Raw material | COMPANY/RM/YY/ABCD | RM = raw material |
| Packing material | COMPANY/PM/YY/ABCD | PM = packing material |
| Finished product: tablets | COMPANY/TA/YYABCD | TA = tablets |
| Finished product: pellets | COMPANY/PE/YYABCD | PE = pellets |
| Finished product: hard gelatin capsules | COMPANY/HC/YYABCD | HC = hard gelatin capsules |
| Water | COMPANY/PW or BW or RW-DD/MM/YYABCD | PW = purified water; BW = bore water; RW = raw water; DD/MM = date |
| Rinse / swab (cleaning) sample | COMPANY/SW/YY/ABCD | SW = rinse or swab sample |
| Miscellaneous | COMPANY/MS/YY/ABCD | MS = miscellaneous item |
In-process and intermediate samples
Enter in-process and bulk intermediate samples date-wise and serial-number-wise in the finished product register.
Using the number
Write the AR number on the sample label, the worksheet or electronic record, the raw data (printouts, chromatograms), the report and the certificate of analysis so every record can be matched to the sample.
Records and retention
Maintain AR number registers (see annexures). Keep AR records and associated raw data at least 1 year after the expiry date of the respective product batch (EU GMP Part I Chapter 4 clause 4.11, which also refers to 5 years after QP certification if longer; 21 CFR 211.180(a)), longer if local law or the marketing authorisation requires.
Training and distribution
- Training: as per the training SOP.
- Distribution: as per the document control SOP.
Annexures
- Annexure I: AR number register, pellets
- Annexure II: AR number register, rinse/swab cleaning samples
- Annexure III: AR number register, miscellaneous items
Frequently asked questions
Is there a regulatory AR number format? No. The format is an in-house choice; the requirement is traceability and complete records.
Can an electronic system assign AR numbers? Yes, if it is validated, the numbers are unique and sequential, and audit trails protect the record (see data integrity guidance).
Should the batch number appear on the AR? Yes. Chapter 6 clause 6.17 expects the batch number to be in the test record.
See also
References
- US FDA, 21 CFR 211.194, Laboratory records (eCFR, current).
- European Commission, EudraLex Volume 4, Part I Chapter 6 Quality Control (in operation since 1 October 2014), clauses 6.7, 6.8, 6.10, 6.11, 6.17.
- European Commission, EudraLex Volume 4, Part I Chapter 4 Documentation, clause 4.11.
- US FDA, 21 CFR 211.180(a), General requirements (records retention).
Related articles
SOP on Maintenance and Preparation of Anaerobic Culture Suspension
Example SOP for maintaining Clostridium sporogenes and preparing an anaerobic culture suspension that delivers 10 to 100 CFU per inoculum for microbial limit testing.
SOP for Certificate of Analysis (CoA) of Finished Product in Word
Example SOP for preparing a finished product Certificate of Analysis (CoA) from a controlled Microsoft Word template, with data-integrity controls and review.
SOP for Cleaning and Numbering Containers and SS Accessories
Example SOP for identification numbering and cleaning of stainless steel and HDPE containers, scoops, ladles and tubs used in a manufacturing area.
Carton Weighing Balance Cleaning SOP (Packaging)
Example SOP for cleaning a carton weighing balance on a pharma packing line: responsibilities, frequency, cleaning steps, verification and logbook entries.